Size range: 0.3μm~10μm up to 12 channels
Real-time, continuous measurement of particle distribution and concentration
Built-in sensors for temperature and humidity monitoring.
Built-in flow control system to ensure measurement accuracy.
ModBus RTU and MQTT communications for remote online monitoring.
Working Principle | Light scattering |
Particle Concentration Output | PM1.0, PM2.5 and PM10 at μg/m³ |
Particle Quantity Output | 12 channels of>0.3μm, >0.4μm, >0.5μm, >0.6μm, >0.8μm, >1.0μm, >1.5μm, >2.0μm, >2.5μm, >5.0μm, >8.0μm and >10μm at pcs/L |
Measurement Range (size) | 0.3μm to 10μm |
Measurement Range (mass) | 0~200 μg/m³ |
Accuracy(mass) | PM1.0/PM2.5: 0~50μg/m³: ±5μg/m³ 50~100μg/m³: ±10% of reading PM10: 0~50μg/m³: ±7μg/m³: 50~200μg/m³: ±15% of reading |
Volume Flow | 6LPM |
Interface | RS485 and RJ45 |
Communication | Modbus-RTU, MQTT |
Operating Voltage | DC 24V±15% |
Working Condition | 0°C~45°C,0~95%RH (Non-condensing) |
Storage Condition | -20°C~60°C,0-95%RH (Non-condensing) |
Dimension | 217*124*87 mm (Main body dimension) |